Development and validation of a new and economical stability indicating RP-HPLC method for cefixime trihydrate

Authors

  • Ceema Mathew Jawaharlal Nehru Technological University
  • Bellamkonda Swathi Osmania University; Gokaraju Rangaraju College of Pharmacy
  • Makula Ajitha Jawaharlal Nehru Technological University
  • Puvvadi Sathesh Babu Osmania University; Gokaraju Rangaraju College of Pharmacy

DOI:

https://doi.org/10.1590/S1984-82502016000100010

Abstract

The present work describes the development of a new high performance liquid chromatographic (HPLC) method for the determination of Cefixime trihydrate under different stress conditons as specified by ICH. For the analysis, a Phenomenex (250 x 4.6 mm, 5 µm particle size) ODS column and a SPD 20 A UV detector at 289 nm was used. The selected mobile phase was 10 mM disodium hydrogen phosphate (with 0.5% TEA, pH adjusted to 6.3 with OPA) and methanol in the ratio of 75:25 (v/v) in isocratic mode at a flow rate of 1 mL.min-1.The linear regression analysis data for the calibration plots showed good linear relationship with r2 = 0.9997 in the concentration range of 5-100 μg.mL-1. The stress degradation was performed using acid, alkali, water, hydrogen peroxide and uv light.

Downloads

Download data is not yet available.

Downloads

Published

2016-03-01

Issue

Section

Articles

How to Cite

Development and validation of a new and economical stability indicating RP-HPLC method for cefixime trihydrate . (2016). Brazilian Journal of Pharmaceutical Sciences, 52(1), 87-94. https://doi.org/10.1590/S1984-82502016000100010